Comparación de tres instrumentos de tomografía de coherencia óptica, un time-domain y dos Fourier-domain, en la estimación del grosor de la capa de fibras nerviosas de la retina

  1. B. Sánchez-Dalmau
  2. V. Llorenç
  3. S. Ortiz
  4. L. Martínez
Journal:
Archivos de la Sociedad Española de Oftalmologia

ISSN: 0365-6691

Year of publication: 2010

Volume: 85

Issue: 5

Pages: 165-173

Type: Article

DOI: 10.1016/S0365-6691(10)55002-0 DIALNET GOOGLE SCHOLAR lock_openOpen access editor

More publications in: Archivos de la Sociedad Española de Oftalmologia

Abstract

Purpose: To assess agreement between one system of time-domain optical coherence tomography (OCT) (Stratus) and two systems of fourier-domain OCT (Cirrus and 3D OCT-1000) for the measurement of the retinal nerve fibre layer thickness (RNFL). Methods: Fifty eyes from 25 patients with neuro-ophthalmological disorders of the afferent visual pathway were included. A triple RNFL thickness OCT examination with Stratus, Cirrus and 3D OCT-1000 was performed on both eyes of every patient, on the same day and by the same technician. The average RNFL thickness by quadrants and clock sectors were compared with the Lin´s concordance correlation coefficient (CCC) and Bland-Altman plots. Results: The average RNFL thickness, the temporal and nasal quadrants and corresponding clock sectors, was higher by 3D OCT-1000 (mean 90.02 microns), nevertheless, RNFL thickness of the superior and inferior quadrants and corresponding clock hours was higher when measured by Stratus device. The CCC agreement between Stratus and Cirrus was high (0.820), between Stratus and 3D OCT-1000 was moderate (mean 0.573) and between Cirrus and 3D OCT-1000 was moderate with a mean of 0.564 but too low in the nasal and temporal quadrants (0.362 and 0.347 respectively). Conclusions: The RNFL thickness measurements by Stratus, Cirrus and 3D OCT-1000 OCT are not equivalent. Although the agreement between Stratus and Cirrus was high, it was too low between 3D OCT-1000 and Stratus or Cirrus.

Bibliographic References

  • Kallenbach, K, Frederiksen, J. (2007). Optical coherence tomography in optic neuritis and multiple sclerosis: a review. Eur J Neurol.. 14. 841
  • Amat Peral, P, Lugo Quintás, FL, de la Vega Galiana, C, Montero Moreno, JA, Ruiz Moreno, JM. Descripción de nuevos sistemas de tomografía de coherencia óptica: Actualizaciones tecnológicas en oftalmología.
  • Blumenthal, EZ, Parikh, RS, Pe´er, J, Naik, M, Kaliner, E, Cohen, MJ. (2009). Retinal nerve fibre layer imaging compared with histological measurements in a human eye. Eye. 23. 171
  • Toledo, J, Sepulcre, J, Salinas-Alaman, A, García-Layana, A, Murie-Fernández, M, Bejarano, B. (2008). Retinal nerve fiber layer atrophy is associated with physical and cognitive disability in multiple sclerosis. Mult Scler.. 14. 906
  • Noval, S, Contreras, I, Rebolleda, G, Muñoz-Negrete, FJ. (2006). Optical coherence tomography versus automated perimetry for follow-up of optic neuritis. Acta Ophthalmol Scand.. 84. 790
  • Vessani, RM, Moritz, R, Batis, L, Zagui, RB, Bernardoni, S, Susanna, R. (2009). Comparison of quantitative imaging devices and subjective optic nerve head assessment by general ophthalmologists to differentiate normal from glaucomatous eyes. J Glaucoma. 18. 253
  • Sung, KR, Wollstein, G, Bilonick, RA, Townsend, KA, Ishikawa, H, Kagemann, L. (2009). Effects of age on optical coherence tomography measurements of healthy retinal nerve fiber layer, macula, and optic nerve head. Ophthalmology. 116. 1119
  • Budenz, DL, Fredette, MJ, Feuer, WJ, Anderson, DR. (2008). Reproducibility of peripapillary retinal nerve fiber thickness measurements with stratus OCT in glaucomatous eyes. Ophthalmology. 115. 661
  • Cettomai, D, Pulicken, M, Gordon-Lipkin, E, Salter, A, Frohman, TC, Conger, A. (2008). Reproducibility of optical coherence tomography in multiple sclerosis. Arch Neurol.. 65. 1218
  • Vizzeri, G, Weinreb, RN, González-García, AO, Bowd, C, Medeiros, FA, Sample, PA. (2009). Agreement between Spectral-domain and Time-domain OCT for measuring RNFL thickness. Br J Ophthalmol.. 93. 775
  • Monteiro, ML, Cunha, LP, Vessani, RM. (2008). Comparison of retinal nerve fiber layer measurements using Stratus OCT fast and regular scan protocols in eyes with band atrophy of the optic nerve and normal controls. Arq Bras Oftalmol.. 71. 534
  • Ghadiali, Q, Hood, DC, Lee, C, Manns, J, Llinas, A, Grover, LK. (2008). An analysis of normal variations in retinal nerve fiber layer thickness profiles measured with optical coherence tomography. J Glaucoma. 17. 333
  • Knight, OJ, Chang, RT, Feuer, WJ, Budenz, DL. (2009). Comparison of retinal nerve fiber layer measurements using time domain and spectral domain optical coherent tomography. Ophthalmology. 116. 1271
  • Hood, DC, Raza, AS, Kay, KY, Sandler, SF, Xin, D, Ritch, R. (2009). A comparison of retinal nerve fiber layer (RNFL) thickness obtained with frequency and time domain optical coherence tomography (OCT). Opt Express. 17. 3997-4003
  • Krebs, I, Falkner-Radler, C, Hagen, S, Haas, P, Brannath, W, Lie, S. (2009). Quality of the threshold algorithm in age-related macular degeneration: Stratus versus Cirrus OCT. Invest Ophthalmol Vis Sci.. 50. 995-1000
  • Kiernan, DF, Hariprasad, SM, Chin, EK, Kiernan, CL, Rago, J, Mieler, WF. (2009). Prospective comparison of cirrus and stratus optical coherence tomography for quantifying retinal thickness. Am J Ophthalmol.. 147. 267
  • Kakinoki, M, Sawada, O, Sawada, T, Kawamura, H, Ohji, M. (2009). Comparison of macular thickness between Cirrus HD-OCT and Stratus OCT. Ophthalmic Surg Lasers Imaging.. 40. 135
  • Costa-Cunha, LV, Cunha, LP, Malta, RF, Monteiro, ML. (2009). Comparison of Fourier-domain and time-domain optical coherence tomography in the detection of band atrophy of the optic nerve. Am J Ophthalmol.. 147. 56-63
  • Wu, Z, Vazeen, M, Varma, R, Chopra, V, Walsh, AC, LaBree, LD. (2007). Factors associated with variability in retinal nerve fiber layer thickness measurements obtained by optical coherence tomography. Ophthalmology. 114. 1505
  • Cheung, CY, Leung, CK, Lin, D, Pang, CP, Lam, DS. (2008). Relationship between retinal nerve fiber layer measurement and signal strength in optical coherence tomography. Ophthalmology. 115. 1347
  • Zheng, Y, Cheung, N, Aung, T, Mitchell, P, He, M, Wong, TY. (2009). Relationship of retinal vascular caliber with retinal nerve fiber layer thickness: The Singapore Malay eye Study. Invest Ophthalmol Vis Sci.. 50. 4091